Research Catalog

Determination of total X-ray beam energy with a calibrated ionization chamber

Title
Determination of total X-ray beam energy with a calibrated ionization chamber / John S. Pruitt and Steve R. Domen.
Author
Pruitt, John S.
Publication
Washington D.C. : U.S. Dept. of Commerce, National Bureau of Standards ; 1962.

Items in the Library & Off-site

Filter by

1 Item

StatusFormatAccessCall NumberItem Location
TextUse in library 8294.739Off-site

Details

Additional Authors
  • Domen, Steve R.
  • United States. National Bureau of Standards.
Description
ii, 16 pages : illustrations, diagrams, tables; 27 cm.
Summary
This report describes the use of an air-filled aluminum-alloy ionization chamber to determine the energy transported by a bremsstrahlung beam with maximum photon energy in the range 6 to 170 Mev. The experimental calibrations of this chamber over this energy range are given, as well as the results of calibration experiments made with a 250-kv constant-potential X-ray tube and with Cs137 and Co60 gamma-rays. Information is presented about the change in calibration when the chamber is used with different experimental conditions, and when either its dimensions or its alloy composition are changed slightly. This report can be used to construct a replica chamber and to determine its absolute calibration between 6 and 170 Mev under a variety of experimental conditions.
Series Statement
National Bureau of Standards Monograph ; 48
Uniform Title
NBS monograph ; 48.
Subject
  • X-rays
  • Ionization chambers
  • x-ray (radiation)
  • Ionization chambers
  • X-rays
  • X rays
Genre/Form
  • technical reports.
  • Technical reports.
  • Rapports techniques.
Note
  • "Issued June 5, 1962."
Bibliography (note)
  • Bibliography: p. 16.
Additional Formats (note)
  • Also available in an electronic version.
LCCN
62061618
OCLC
  • ocm04915913
  • 4915913
  • SCSB-306117
Owning Institutions
Princeton University Library